Set the multi-wavelength UV detector to measure absorbance at wavelengths (λ) of 280 nm and 260 nm during the run. The relative A280/A260 ratio of the AAV particles is used during analysis to indicate the full and empty AAV capsids. For more detail, see the example protocol displayed in Table 1 below.
Column Performance (Optional):
Attach the column to your chromatography system and wash it with water to remove the storage solution. Next, prime the column with the equilibration buffer A, inject the tracer (e.g., 1 M NaCl), and then chase with the same equilibration buffer. Measure the asymmetry and HETP to evaluate the column performance.
Equilibration:
Equilibrate column by passing buffer A, the paired buffer B, and then buffer A again. Allow the column to reach pH and conductivity stability (i.e., signal plateau) respective of the buffer attributes at each of the three steps.
Sample Application:
Apply the sample. Wash column with buffer A until absorbances reach baseline or plateau, ensuring any loosely bound material is fully passed through. At a decreased flow rate, elute AAV using a linear gradient from 0% to 50% of buffer B over 50 column volumes. Collect eluate into fixed volume fractions.
- TIP: Smaller fraction volumes will allow a more precise collection of the “full” AAV samples.
Strip column with 100% of buffer B to remove more tightly bound material. For further column cleaning and/or regeneration, follow recommended protocols for your column.
Column Regeneration:
Wash column with water and follow with a CIP solution, water, 1 M Ammonium Acetate to neutralize column pH, water, and then re-equilibrate with buffer A. The column can then either be stored in vendor recommended storage buffer or solution (e.g., 20% Ethanol), or the process can be restarted from column performance if additional purification of sample(s) is needed.